Data sheet 10/23-5.20 EN Rev. 13

System components for sample gas conditioning
Sample gas cooler, gas feed unit, converter, accessories

Measurement made easy

SCC-C Sample gas cooler

Acid filter

SCC-F Sample gas feed unit

Disposable filter

SCC-S Sample gas feed unit

Flow meters, flow monitors

Diaphragm pump 4N

ER-144A/Ex Switching unit

SCC-K NO2 / NO converter

KFA Isolated switch amplifier

CGO-9 Tubular furnace

Stripper for cooling water and waste water

Reaction element

Multi-way test gas cocks

Membrane filter

Cylinder pressure reducers

CGKW Condensate monitor

SCC-C Sample gas cooler
 Compressor sample gas cooler
 Constant outlet dew point
 1 or 2 gas paths
 Heat exchanger made from glass, PVDF or stainless steel,
exchangeable without tools
 Corrosion-resistant gas lines
 High pressure design up to 1 MPa (10 bar)
 Temperature display
 Suitable for ambient temperatures up to 50 °C
 Compact surface-mounting housing (width 1/2 19 inches,
6 height units)
 System solutions in connection with the SCC-F sample
gas feed unit
 Suitable for use in measuring equipment conforming with
17th BImSchV (German Federal Immission Protection Law)
 Explosion-proof versions with ATEX and CSA
certifications

Sample gas inlet conditions

Dead volume in heat exchanger
Heat exchanger material

Sample gas pressure
Heat exchanger Sample gas pressure pabs
(HE) material
w/o peristaltic pump with peristaltic pump
Glass
50–200 kPa
50–150 kPa
(0.5–2.0 bar)
(0.5–1.5 bar)
PVDF
50–250 kPa
50–150 kPa
(0.5–2.5 bar)
(0.5–1.5 bar)
Stainless steel 0.05–1 MPa
50–150 kPa
(0.5–10 bar)
(0.5–1.5 bar)

Glass
PVDF
Stainless steel

Dead volume in heat exchanger
HE125
HE250
40 ml
140 ml
25 ml
100 ml
30 ml
100 ml

Gas tightness
5 x 10–6 hPa l/s

Display and status signal

Sample gas flow rate
1 x 250 l/h (HE250) or 1 x 125 l/h (HE125) or 2 x 125 l/h, relating to sample gas pressure pabs = 100 kPa (1 bar) and 25 °C

Temperature display
Digital display in °C

Sample gas inlet temperature
Max. 140 °C

Status signal
< 0 °C and > 8 °C, 2 potential-free two-way switches,
switching capacity 250 V AC, 2 A

Sample gas inlet dew point
Max. 70 °C,
max. 60 °C for HE250 at sample gas flow rate > 200 l/h

Construction
Design
Surface-mounting housing 1/2 19 inch, 6 height units

Operating data
Sample gas outlet temperature
Factory-set to +3 °C

Mounting
On the wall with mounting brackets supplied or
in a 19-inch cabinet / rack

Dew point stability
0.3 °C per 10 °C temperature change,
0.3 °C per 10 l/h change in flow rate

Housing material and color
Sheet steel; light gray (RAL 7035), basalt gray (RAL 7012)

Refrigeration capacity
40 W at +10 to +50 °C

Dimensions
See dimensional drawing on page 4

Warm-up time
Approx. 15 min
Pressure drop in heat exchanger
Approx. 1 hPa (1 mbar), approx. 4–8 hPa (4–8 mbar) for
HE125

2

Degree of protection of housing
IP20 to EN 60529

Weight
Approx. 15.6 / 16.5 kg with 1/ 2 heat exchangers
Refrigerant
R 134a

System components and accessories for sample gas conditioning

10/23-5.20 EN Rev. 13

SCC-C Sample gas cooler
Materials of gas-conducting parts

Ambient conditions

PVDF, PTFE, PP, glass, FPM, stainless steel (high pressure
version)

Ambient temperature
Operation: +5 to +45 °C,
Storage and transport: –25 to +60 °C

Gas connections

Relative humidity
 75 % annual average, occasional and slight condensation
permitted,  95 % on 30 days per year

Heat exchanger
Heat exchanger Sample gas inlets &
material
outlets
Glass
PVDF
Stainless steel

Condensate outlet

Screwed pipe or hose Pipe nipple GL25
connection GL18 1)
6 mm tube
G 3/8 female thread
G 1/4 female thread
G 3/8 female thread

1) 2 GL coupling nuts per heat exchanger with insert 6 x 4 x
1 mm threads for a hose or pipe are supplied with the unit
Peristaltic pump (option)
Condensate outlet PVDF, DN 4/6

Power supply
Input voltage
230 V AC or 115 V AC –15 to +10 %, 50 to 60 Hz
Power consumption
Max. 200 VA
Starting current
2.5 A at 230 V AC

Options
Electrical connections
Power supply
Cable, 3 m long, permanently fixed
Status signal
Cable, 3 m long, permanently fixed

Electrical safety
Testing to EN 61010-1
Protective class
I
Overvoltage category / degree of contamination
III / 2
Protective separation
Electrical isolation of the 115 / 230 V AC power supply from
the other current circuits by means of reinforced or double
insulation. Functional extra-low voltages (PELV) on the low
voltage side

Metering device
In gas path 1 by means of a peristaltic pump installed in the
SCC-C sample gas cooler or in the SCC-F sample gas feed
unit, feed tank as accessory, for small SO2 measuring ranges
(< 1000 mg/m3) or to avoid acid aerosols
Peristaltic pumps
1 or 2 peristaltic pumps for discharging condensate
(flow rate approx. 300 ml/h, 5 rpm) and metering (flow rate
approx. 15 ml/h),
Pressure resistance pabs = 50–150 kPa (0.5–1.5 bar),
Peristaltic tube service life approx. 5 months,
Power consumption 3.5 VA,
Weight approx. 0.6 kg

Explosion-proof versions
ATEX certification
The SCC-C sample gas cooler in the category 3G explosionproof version is suitable for use in zone 2 hazardous areas.
Testing to EN 60079-15 : 2003

Electromagnetic compatibility
Emitted interference
Testing to EN 61000 Part 6-3 : 2001
Immunity to interference
Testing to EN 61000 Part 6-2 : 2005

Mechanical capabilities
Testing to EN 60068-2-27 and EN 60068-2-6
Transport
Shock: 30 g / 18 ms / 18 shocks
Vibration: 1 g / ±0.15 mm / 5–150 Hz, 3 x 5 cycles
Operation
Vibration: 1 g / ±0.07 mm / 5–55 Hz, 3 x 2 cycles

10/23-5.20 EN Rev. 13

Designation:

II 3G Ex nAC [nL] IIC T4

Type examination certificate: KEMA 06 ATEX 0165X
CSA certification
The SCC-C sample gas cooler is certified to
– Class 2258 02 Process Control Equipment – For Hazardous
Locations and
– Class 2258 82 Process Control Equipment – For Hazardous
Locations – Certified to U.S. Standards
Class 1, Div. 2, Groups A, B, C and D, Temperature Code T4,
Maximum Ambient Temperature +50 °C.
Approval includes testing per applicable Canadian (CSA) and
US directives.
Certificate No. 1105720

System components and accessories for sample gas conditioning

3

SCC-C Sample gas cooler
Dimensional drawing (dimensions in mm)

13

1
2

266

1

3
4

190.5

3.0

Temperature display
Condensate outlet (for option with
peristaltic pumps)
Gas connections of heat exchangers
Conduits for the (permanentlyconnected) electrical connection lines

37.8

2

246
263
220

326.5

4

46

3

Circuit and connection diagram
Bridge = 230V Transformer = 115V

Mains L

F1

X1/4

X2/2
L

2A T

X2/3

275V

R2

L

X4/3

X4/2

M
~

X5/O/5
M1

1
Pt 100

X4/1

Option
Pt 100

X5/O/3

X2/1

U

T

T

R1

3

X4/4

C

R

S
X5/O/2

5+

X6/4

X3/3

X5/M/2

Solid

X3/1

6-

C1
4,7μF

A1
X6/3

X5/O/1

X5/U/2
X5/M/3

K1

Alarm

Status 1
X3/2

8

X6/2

9

X6/1

X5/O/4
Fan
M2

X3/6
N
X3/4

MV1

K1

X5/O/7

SR25
M4
M
~

X5/O/6

SR25
M3
M
~

M
~

A2
K1

Status 2

X5/M/5

X5/M/1

X5/M/4

X5/M/7

X5/M/6

X3/5

Mains N
X1/3

Mains PE
X1/2

X5/U/1

X5/U/4

X5/U/3

X1/1

4

System components and accessories for sample gas conditioning

10/23-5.20 EN Rev. 13

SCC-C Sample gas cooler
Refrigerant circuit
1
2
3
4
5
6
7
8
9

2
1

8

Refrigerant compressor
Refrigerant condenser
Refrigerant dryer
Capillary tube
Heat exchanger
Evaporator
Cooling block with temperature sensor
Valve
Temperature controller

3

4
9

°C

7

5
6

Accessories for condensate disposal
Time relay

Level monitor

Run time and off period settings
0.6–60 sec., recommended setting: cyclic duration factor
40 %, i.e. run time 5 minutes and off period 7.5 minutes

Switching voltage
30 V

Degree of protection
Terminals IP20, housing IP50 to EN 60529
Weight
approx. 0.125 kg

Switching current
1A
Switching power
30 VA
Connecting cable
2 x 0.25 mm2, length 5 m

Power supply
12–240 V AC/DC, 40–440 Hz

Ordering information
Condensate collecting bottle
Volume
10 l
Material
LDPE / PVC
Connections
Sample gas: 1 connecting piece for hose with ID 4–5 mm
Condensate: 2 connecting pieces for hose with ID 12–13 mm
Condensate drain: SW 22

Timer relay
Reagent and condensate bottle
Level monitor condensate
Level monitor reagent
Condensate trap, filter surface 20 cm2

Catalog No.
90805-4-0741607
23234-5-8328972
23005-4-0741239
90805-5-8308994
23065-5-8018512

Weight
approx. 1 kg

10/23-5.20 EN Rev. 13

System components and accessories for sample gas conditioning

5

SCC-C Sample gas cooler
Ordering information
Catalog No. 23070-0
Design
Wall-mounted housing 1/2 19 Inch













0


0


0



0


0


0


0

1

Preparation
without
with dosing unit

0
1

Heat exchanger
Glass
PVDF
Stainless steel

1
2
3

Gas paths
1 gas path 250 l/h
1 gas path 125 l/h
2 gas paths 2 x 125 l/h

1
2
3

Condensate drain
without
1 peristaltic pump
2 peristaltic pumps (2 x condensate)
2 peristaltic pumps (1 x condensate, 1 x dosing)
2 peristaltic pumps (1 x condensate, 1 x precondensation)

0
1
2
3
4

Display
Temperature display and monitoring
Temperature display and monitoring and Pt 100

1
2

Power supply
230 V, 50/60 Hz
115 V, 50/60 Hz

1
2

Certification
without
with CSA certification
with ATEX certification, Zone 2
with GOST certification

0
1
2
3






Catalog No. 23070-0-

6



0


System components and accessories for sample gas conditioning











10/23-5.20 EN Rev. 13

SCC-F Sample gas feed unit
 Complete sample gas conditioning in connection with the
SCC-C sample gas cooler
 2 independent gas paths
 Corrosion resistant, metal-free gas lines
 Compact surface-mounting housing (width 1/2 19 inches,
6 height units)
 Service-friendly due to easy access to the modules

Sample gas inlet conditions

Feed performance standard
250

Sample gas pressure
pabs = 70–105 kPa (0.7–1.05 bar)

Sample gas inlet temperature
+10 to +50 °C
Sample gas inlet dew point
The dew point of the sample gas must be at least 5 °C lower
than the lowest ambient temperature in the whole sample
gas path. Otherwise a sample gas cooler or a condensate
trap is required.

200
150

100
50

-800

-600

Inlet pressure
0

Flow rate / l/h

Sample gas flow rate
Gas paths 1 and 2 max. 100 l/h, bypass max. 200 l/h

A

-400

B

hP
a

-1
00
h
-2 Pa
00
-3 hP
00 a
hP
a
-4
00
hP
a

-200
0
200
Pressure / hPa

400

600

800

A Inlet: negative pressure, outlet: atmospheric pressure
B Inlet: acc. to chart, outlet: positive pressure

Operating data
Warm-up time
Approx. 10 minutes (plus warm-up time of sample gas
cooler)

Feed performance bypass
350

Dead volume
approx. 10 cm3 (plus dead volume of heat exchanger)

300

Flow rate display and adjustment
Rotameter 10–100 l/h with needle valve (metal-free)

Flow / l/h

250

Gas tightness
5 x 10–6 hPa l/s

200
150
100
50

-1000

10/23-5.20 EN Rev. 13

-500
0
500
Pressure / hPa

System components and accessories for sample gas conditioning

1000

1500

2000

2500

7

SCC-F Sample gas feed unit
Status signals

Mechanical capabilities

Status signals
Flow rate and condensate monitors: 2 potential-free
two-way switches, switching capacity 250 V AC, 2 A

Testing to EN 60068-2-27 and EN 60068-2-6

Construction

Operation
Vibration: 1 g / ±0.07 mm / 5–55 Hz, 3 x 2 cycles

Design
Surface-mounting housing 1/2 19 inch, 6 height units
Mounting
On the wall with the mounting brackets supplied or
in a 19 inch cabinet / rack
Housing material and color
Sheet steel; light gray (RAL 7035), basalt gray (RAL 7012)
Degree of protection of housing
IP20 to EN 60529
Dimensions
See dimensional drawing on page 9

Material of gas-conducting parts
PVDF, PTFE, PVC, EPDM, FPM, PP-EPDM elastomer

Transport
Shock: 30 g / 18 ms / 18 shocks
Vibration: 1 g / ±0.15 mm / 5–150 Hz, 3 x 5 cycles

Ambient conditions
Ambient temperature
Operation: +5 to +45 °C,
Storage and transport: –25 to +60 °C
Relative humidity
 75 % annual average, occasional and slight condensation
permitted,  90 % on 30 days/year

Power supply
Input voltage
230 V (207–250 V) AC, 50 Hz or 60 Hz  2 Hz
115 V (100–135 V) AC, 50 Hz or 60 Hz  2 Hz
Power consumption during operation
Diaphragm pump approx. 50 VA,
metering pump approx. 3.5 VA

Gas connections
PVDF coupling for tube of 6 x 4 x 1 mm

Options
Electrical connections
Power supply
Screw clamp terminals for stranded or solid wire with
max. 1.5 mm2
Status signals
Screw clamp terminals for stranded or solid wire with
max. 1.5 mm2

Electrical safety
Testing to EN 61010-1
Protective class
I
Overvoltage category / degree of contamination
III / 2
Protective separation
Electrical isolation of the 115 / 230 V AC power supply from
the other current circuits by means of reinforced or double
insulation. Functional extra-low voltages (PELV) on the low
voltage side

Electromagnetic compatibility
Emitted interference
Testing to EN 61000 Part 6-3 : 2001
Immunity to interference
Testing to EN 61000 Part 6-2 : 2005

8

Metering unit
By means of peristaltic pump, flow rate 15 ml/h,
feed tank as accessory
for small SO2 measuring ranges (< 1000 mg/m3) or
to avoid acid aerosols
I/O card
To connect the SCC-F sample gas feed unit and the SCC-C
sample gas cooler to the AO2000 Series continuous gas
analyzers via the system bus
– Display and monitoring of the cooler temperature
– Monitoring and processing of status signals of the SCC-F
sample gas feed unit (condensate, flow rate)
– Control of an external solenoid valve for zero gas
connection, 230 V, 50 Hz
– Digital inputs for monitoring condensate collecting bottles
and reagent reservoirs

CSA certification
The SCC-F sample gas feed unit is certified to
– Class 2258 02 Process Control Equipment – For Hazardous
Locations and
– Class 2258 82 Process Control Equipment – For Hazardous
Locations – Certified to U.S. Standards
Class 1, Div. 2, Groups A, B, C and D, Temperature Code T4,
Maximum Ambient Temperature +50 °C.
Approval includes testing per applicable Canadian (CSA) and
US directives.
Certificate No. 1105720

System components and accessories for sample gas conditioning

10/23-5.20 EN Rev. 13

SCC-F Sample gas feed unit
Dimensional drawing (dimensions in mm)
Z

Ansicht / View Z

20

220

2

Sample gas inlet
Sample gas outlet
Electrical connections
Flow rate monitor with needle valve
Metering pump
Condensate monitor
LED display and reset button of
switching amplifiers

6
7
326.5

266

190.5

6

1

1
2
3
4
5
6
7

37.8

4

4
5
246

46

263

Ordering information
Catalog No. 23212-0
Design
Wall-mounted housing 1/2 19 Inch









0


0


0


0


0



0


0


0


0


0


0

1

Gas paths
1 gas path, 1 condensate monitor, 1 flow monitor, 1 diaphragm pump
2 separate gas paths, 2 condensate monitors, 2 flow monitors, 2 diaphragm pumps
1 gas path, 1 condensate monitor, 2 flow monitors, 1 diaphragm pump
1 gas path, 1 condens. monitor, 2 flow monitors, 1 diaphr. pump, 1 pressure controller
2 separate gas paths, 2 condensate monitors, 2 flow monitors, 1 diaphragm pump
2 separate gas paths, 2 condensate monitors, 1 flow monitor, 1 diaphragm pump

1
2
3
4
5
6

Dosing
without
with dosing pump

0
1

Electronic processing unit
with 1 electronic processing unit
with 2 electronic processing units
with I/O interface board

1
2
3

Power supply
230 V, 50/60 Hz
115 V, 50/60 Hz

1
2

Certification
without
with CSA certification
with GOST certification

0
1
2






Catalog No. 23212-0-

10/23-5.20 EN Rev. 13



0


System components and accessories for sample gas conditioning







9

SCC-F Sample gas feed unit
Pneumatic diagrams
Catalog number 23212-0-11XXXX000000

Catalog number 23212-0-12XXXX000000

1

1

1

1

2

2

1

2

Catalog number 23212-0-14XXXX000000

1

Catalog number 23212-0-15XXXX000000

1
1

10

Catalog number 23212-0-16XXXX000000

1

2

1

Catalog number 23212-0-13XXXX000000

1
2

2

1
2

2

System components and accessories for sample gas conditioning

10/23-5.20 EN Rev. 13

SCC-S Sample gas feed unit
 Process-proven modules
– Diaphragm pump for feed-in of the gas
– Membrane filter/condensate monitor
– Flow monitor with needle valve
– Ball valve/solenoid valves for feed-in of the test gas
– Peltier cooler with condensate pump
 Ideal for use in simple processes (non-corrosive sample
gases)
 Simple installation and connection
 Housing designed for 19-inch rack mounting and wall
mounting (4 height units, IP20)
 Attractively priced system solution in combination with
gas analyzers of the EasyLine series

Sample gas inlet conditions

Feed performance of the diaphragm pump
250

Sample gas pressure
pabs = 70–105 kPa (0.7–1.05 bar)

Sample gas inlet temperature
+10 to +50 °C
Sample gas inlet dew point
Versions with Peltier cooler: max. +50 °C (at max. +25 °C
ambient temperature)
Versions without Peltier cooler: The dew point of the sample
gas must be at least 5 °C lower than the lowest ambient
temperature in the whole sample gas path.

150

100
50

-800

-600

Inlet pressure
0

200

Flow rate / l/h

Sample gas flow rate
Gas path: 7–70 l/h,
Bypass: 15–150 l/h (not in versions with Peltier cooler)

A

-400

B

hP
a

-1
00

h
-2 Pa
00
h
-3 P
00 a
hP
a
-4
00
hP
a

-200
0
200
Pressure / hPa

400

600

800

A Inlet: negative pressure, outlet: atmospheric pressure
B Inlet: acc. to diagram, outlet: positive pressure

Operating data
Warm-up time
Versions with Peltier cooler: approx. 10 minutes
Versions without Peltier cooler: ready for immediate use
Gas tightness
5 x 10–6 hPa l/s
Flow rate display and adjustment
Rotameter with needle valve (metal-free)
Outlet dew point (versions with Peltier cooler)
+3 °C
Condensate discharge (versions with Peltier cooler)
Peristaltic pump: feed performance approx. 300 ml/h at
5 rpm, pressure resistance pabs = 50–150 kPa (0.5–1.5 bar),
peristaltic tube material TPE, peristaltic tube service life
approx. 5 months, power consumption 3.5 VA, weight
approx. 0.6 kg

10/23-5.20 EN Rev. 13

Display and status signals
Temperature control and display (versions with Peltier cooler)
Temperature controller with digital display of the cooler
temperature in °C; set point factory-set at +3 °C, can be
altered on site.
Status signals
Collective status signal of the condensate, flow and cooler
monitoring: potential-free changeover contact, contact load
24 V DC/AC, 1 A
Status display of the condensate monitoring: Red LED on
the front panel

System components and accessories for sample gas conditioning

11

SCC-S Sample gas feed unit
Applications: Peltier cooler for combustion gases

Material of gas-conducting parts

Application 1
SO2 0–75 mg/m3, NO 0–100 mg/m3, measurement with
Uras26

Glass, glass fiber, EPDM, ETFE, FPM, PP, PPH, PVC, PVDF,
TPE

Sample gas inlet dew
point at the cooler inlet
30 °C

50 °C

Ambient
Flow rate
temperature 30 l/h 60 l/h
X
X
+10 °C
+20 °C
X
X
+30 °C
X
X
+40 °C


+10 °C
X
X
+20 °C
X
X
+30 °C
X

+40 °C



90 l/h
X



X




X: permitted, –: not permitted
Application 2
SO2 0–400 mg/m3, NO 0–100 mg/m3, measurement with
Uras26 (also for CO, CO2 and O2)
Sample gas inlet dew
point at the cooler inlet
30 °C

50 °C

Ambient
Flow rate
temperature 30 l/h 60 l/h
X
X
+10 °C
+20 °C
X
X
+30 °C
X
X
+40 °C
X
X
+10 °C
X
X
+20 °C
X
X
+30 °C
X
X
+40 °C
X
X

90 l/h
X
X
X
X
X
X
X
X

Gas connections
Sample gas inlet
PVDF angled screw fitting for tube 6 x 4 x 1 mm
Test gas inlets
PPH coupling for tube 6 x 4 x 1 mm
Gas outlets
PPH coupling for tube 6 x 4 x 1 mm

Electrical connections
Power supply
Connector for non-heating apparatus (cable with grounding
plug, length 2 m, supplied)
Status signal (standard) and external solenoid valves (option)
8-pin plug receptacle (see connection diagram, mating plug
supplied)

Connection diagram

Construction
Design
19-inch housing (4 height units) for rack and wall mounting
Mounting
Rack mounting on mounting rails, wall mounting with preassembled mounting brackets on the housing.
The gas and electrical connections are located on the back
of the housing for rack mounting and on the top of the
housing for wall mounting. They can be altered on site.
Comply with the distances from the wall during installation:
left and right-hand side panels at least 35 mm, back panel at
least 25 mm.
A distance of at least 1 height unit from devices which
develop heat is required for the version with Peltier cooler.
Max. inclination of the instrument 5°.
Housing material and color
Sheet steel, galvanized, basalt gray (RAL 7012),
front panel light gray (RAL 7035)
Degree of protection of housing
IP20 to EN 60529
Dimensions
See dimensional drawing on page 15
Weight
Approx. 17 kg

12

1–3 Status signal of the cooler, flow and condensate
monitoring: potential-free changeover contact, contact
load 24 V DC/AC, 1 A
4
Activation of solenoid valve Sample/Test gas
5
Activation of solenoid valve Test gas 1
6
Activation of solenoid valve Test gas 2
7
Internal power supply 0 V
8
Internal power supply +24 V DC
The connection diagram shows the internal wiring of the
sample gas feed unit.
The integrated power supply can also be used for the supply of
external solenoid valves: Terminal 7 = 0 V, terminal 8 = +24 V,
maximum capacity 30 VA at 24 V DC.

System components and accessories for sample gas conditioning

10/23-5.20 EN Rev. 13

SCC-S Sample gas feed unit
Electrical safety

Options

Testing to EN 61010-1

Internal power supply/ power stage
For the activation of max. 3 internal or external solenoid
valves, operating voltage 24 V DC, capacity max. 30 VA, 3 A
fine-wire fuse

Protective class
I
Overvoltage category/degree of contamination
III / 2
Protective separation
Electrical isolation of the 115 / 230 V AC power supply from
the other current circuits by means of reinforced or double
insulation. Functional extra-low voltages (PELV) on the low
voltage side

Electromagnetic compatibility
Emitted interference
Testing to EN 61000 Part 6-3 : 2001
Immunity to interference
Testing to EN 61000 Part 6-2 : 2005

Solenoid valves for internal test gas infeed
3/2-way solenoid valve: 1 ea., for switching over sample
gas/test gas, valve type is pivoted armature valve, material
PVDF and FPM, power supply 24 V DC, activation via a
digital output of the gas analyzer or a PLC.
2/2-way solenoid valve: max. 2 ea., for test gas infeed, valve
type is rocker valve with separating diaphragm, material
PPH and FPM, power supply 24 V DC, activation via a digital
output of the gas analyzer or a PLC.
Condensate discharge for external cooler
Peristaltic pump: flow rate approx. 300 ml/h at 5 rpm,
pressure resistance pabs = 50–150 kPa (0,5–1,5 bar), peristaltic
tube material TPE, peristaltic tube service life approx.
5 months, power consumption 3.5 VA, weight approx. 0.6 kg

Mechanical capabilities
Testing to EN 60068-2-27 and EN 60068-2-6
Operation
Vibration: 1 g / ±0.08 mm / 10 to 150 Hz, 3 x 2 cycles

Ambient conditions
Ambient temperature
Operation: +10 to +40 °C,
Storage and transport: –25 to +60 °C
Relative humidity
 75 % annual average, occasional and slight condensation
permitted,  90 % on 30 days/year

Power supply
Input voltage
230 V AC ± 10 %, 50 Hz or 115 V AC ± 10 %, 60 Hz
Power consumption during operation
230 V AC, 50 Hz, max. 300 VA
Protection of the instrument
Fine-wire fuse, 3.15 A, slow-blow

10/23-5.20 EN Rev. 13

System components and accessories for sample gas conditioning

13

SCC-S Sample gas feed unit
Pneumatic diagrams
Version 1

Catalog number 23236-0-X11000XX0000

Version 6

Catalog number 23236-0-X2001XXX0000

3

3

2

6
5

1

B

1

6
Version 2

Catalog number 23236-0-X21000XX0000

A
3

2

Version 7

7

8
D

Catalog number 23236-0-X2101XXX0000

3

2

1

Version 3

8

Catalog number 23236-0-X31000XX0000
15-150 l/h

1

4
3

2

A

1

Version 8

8
7

8

D

Catalog number 23236-0-X2201XXX0000

3

2
24VDC

Version 4

Catalog number 23236-0-X22000XX0000

3

2
24VDC

1

1

Version 5

A

8
7

D

8

Catalog number 23236-0-X32000XX0000

24VDC

2

15-150 l/h

4
3

1

14

System components and accessories for sample gas conditioning

10/23-5.20 EN Rev. 13

SCC-S Sample gas feed unit
Pneumatic diagrams
Version 9

Catalog number 23236-0-X2301XXX0000

Version 10 Catalog number 23236-0-X2011XXX0000

3

2
24VDC

24VDC

2

6

1

5

C

1

8

A

1
2
3
4
5
6
7
8

3

6

8

7

8

A

7

D

Sample gas inlet
Test gas inlet
Gas outlet
Bypass outlet
Sample gas inlet for external infeed
Test gas inlet for external infeed
Condensate inlet from the external cooler
Condensate outlet

8

D

Not supplied:
A External cooler
B External multi-way cock
C External solenoid valves
Option:
D Additional condensate pump for external cooler

Dimensional drawing (dimensions in mm)
426

482,6

9

37,75

10

101,5

8
177

194

49

5

6

4

37,75

7
8,5

465,6

8,5

433,5

23,5

1
3

353,5

2

1
2

Sample gas inlet
Sample gas outlet, bypass outlet and test gas
inlets via ball valve or solenoid valve
3
Power supply connection and
electrical connection for status signal, solenoid
valve control and 24 V DC
4
Ball valve for infeed of test gas
5
Flow monitor with needle valve (Option: second
flow monitor for bypass)
6
Peristaltic pump (Option: second peristaltic
pump for external cooler)
7
Membrane filter or condensate monitor
8
LED indicator for condensate alarm, red
9
Pump switch
10 Temperature controller for cooler

42,5

Note: The illustration shows options.The scope and
features of the ordered design may differ.

10/23-5.20 EN Rev. 13

System components and accessories for sample gas conditioning

15

SCC-S Sample gas feed unit
Ordering information
Catalog number 23236-0
Design
19-inch housing for rack mounting, 4 height units, IP20
19-inch housing for wall mounting, 4 height units, IP20













0


0


0



0


0


0


0

1
2

Gas path
Membrane filter, pump and flow monitor with needle valve (Version 1)
Condensate monitor, pump and flow monitor with needle valve (Vers. 2, 4, 6–10)
Condensate monitor, pump and flow monitor 7–70 l/h, bypass 15–150 l/h (Vers. 3, 5)

1
2
3

Internal test gas infeed
without (Versions 6, 10)
One 3-way test gas cock (Versions 1, 2, 3, 7)
One 3/2-way solenoid valve (Versions 4, 5, 8)
One 3/2-way solenoid valve and two 2/2-way solenoid valves (Version 9)

0
1
2
3

Electrical power stage (24 VDC, 30 VA) for external test gas infeed
without (Versions 1–9)
For up to three external solenoid valves (Version 10)

0
1

Cooler and condensate pump
without (Versions 1–5)
with Peltier cooler and condensate pump (Versions 6–10)

0
1

Condensate pump for external cooler
without (Versions 1–5)
with additional condensate pump (Versions 6–10)

0
1

Power supply
230 V, 50 Hz
115 V, 60 Hz

1
2

Approvals
None

1






Catalog number 23236-0-

16



0


System components and accessories for sample gas conditioning











10/23-5.20 EN Rev. 13

Diaphragm pump 4N
 Feed in of the sample gas
 Compact unit with degree of protection of housing IP20
or IP54

Operating data

Construction

Inlet pressure
pabs = 70–105 kPa (0.7–1.5 bar)

Design
Surface-mounting case for wall attachment, motor axis
horizontal

Flow rate
Max. 250 l/h at atmospheric pressure

Housing material and color
Plastic, RAL 7035

Inlet temperature
+5 to +50 °C

Degree of protection of housing
IP20 or IP54 to EN 60529

Inlet dew point
Below room temperature

Class of protection
I

Tightness
Pressure drop 0.1 hPa/min

Dimensions
See dimensional drawing on page 18

Feed performance
250
A

0

Flow rate / l/h

200
150

100
50

Weight
Approx. 1.8 kg

Inlet pressure

B

hP
a

-1
00
h
-2 Pa
00
h
-3 P
00 a
hP
a
-4
00
hP
a

Materials of gas-conducting parts
Valve plate, diaphragm: EPDM (ethylene-propylene);
pump body: PP (polypropylene); hose: FPM;
nozzles: ETFE (ethylene-tetrafluorethylene)

Gas connections
-800

-600

-400

-200
0
200
Pressure / hPa

400

600

A Inlet: negative pressure, outlet: atmospheric pressure
B Inlet: according diagram, outlet: positive pressure

800

G 1/4 female thread for male fittings, nozzles for hose inside
diameter 4 mm are supplied with the unit.

Electrical connections
Terminals via cable connection M20

10/23-5.20 EN Rev. 13

System components and accessories for sample gas conditioning

17

Diaphragm pump 4N
Mechanical capabilities

Dimensional drawing (dimensions in mm)

Testing to DIN EN 60068-2-27 and DIN EN 60068-2-6
Transport
Shock: 30 g / 18 ms / 18 shocks
Vibration: 2 g / 0.15 mm / 5–150 Hz, 3 x 5 cycles

Z

Operation
Vibration: 2 g / 0.07 mm / 4–7 Hz, 3 x 2 cycles
2
1

Ambient conditions

32

32

95

140

Z

Relative humidity
 90 % annual average, condensation permitted

16

Climatic class
3K4 to EN 60721-3-3

~150

ø 4,2
88
120

Ambient temperature
Operation: IP54 +5 to +40 °C, IP20 +5 to +50 °C
Storage and transport: –25 to +65 °C

17

60

148
160

Power supply

6

1

Gas connections

2

Electrical connections

Z-14185

AC voltage
230 V  10 %, 50 Hz or 60 Hz
115 V  10 %, 50 Hz or 60 Hz
Power consumption
approx. 50 VA

Ordering information
Diaphragm pump 4N, IP20; 230V, 50/60 Hz
Diaphragm pump 4N, IP20; 115V, 50/60 Hz
Diaphragm pump 4N, IP54; 230V, 50/60 Hz
Diaphragm pump 4N, IP54; 115V, 50/60 Hz

18

Catalog No.
23134-5-8018545
23134-5-8018546
23134-5-8018547
23134-5-8018548

System components and accessories for sample gas conditioning

10/23-5.20 EN Rev. 13

SCC-K NO2 /NO converter
 Catalytic converter based on carbon-molybdenum
 High catalyst service life
 Catalyst cartridge can be changed without tools
 Option: Version with heated sample gas inlet

Operating data

Construction

Sample gas flow rate
Max. 150 l/h

Design
19 inch slide-in housing, 3 height units

Temperature range
Depending on sample gas flow rate:
30 l/h: 320 °C; 60 l/h: 320 °C; 90 l/h: 340 °C, 150 l/h: 360 °C

Mounting
In 19 inch cabinet / rack or with mounting brackets
(accessories) on the wall, distance above min. 1 height unit

Effectivity
 95 % with new catalyst

Housing material and color
Sheet steel; light gray (RAL 7035), basalt gray (RAL 7012)

Sample gas pressure pabs
 200 kPa (2 bar)

Degree of protection of housing
IP20 to EN 60529

Pressure drop
 2 kPa (20 mbar) at 90 l/h

Dimensions
See dimensional drawing on page 21

Warm-up time
Approx. 30 min

Weight
Approx. 8–9 kg

90% time T90
 10 s at 60 l/h

Materials of gas-conducting parts
Stainless steel 1.4571 (SAE 316Ti), PVDF, PTFE, FPM

Status signal
Status signal for temperature deviation
1 NO potential-free contact,
capacity 24 V DC, 1 A

Gas connections
Sample gas inlet and outlet unheated
G 1/4 female thread
Heated sample gas inlet
Swagelok® screwed connection for pipes with 6 mm inner
diameter

10/23-5.20 EN Rev. 13

System components and accessories for sample gas conditioning

19

SCC-K NO2 /NO converter
Electrical connections

Power supply

Power supply
Connector for non-heating apparatus, cable with grounding
plug, length 2 m

Input voltage
230 V AC, –15 / +10 %, 48–62 Hz or
120 V AC, –10 / +10 %, 48–62 Hz

Status signal
9 pin Sub-D female connector with counter-plug

Power consumption
240 V AC: 575 VA; 120 V AC: max. 560 VA

Electrical safety

Service life of catalyst

Testing to EN 61010-1

The service life of the catalyst depends on its operating temperature as well as the flow rate and the NO2 and O2 content of
the sample gas. It is > 6 months for 320 °C, 30 l/h, 10 ppm NO2
and 5 Vol.-% O2.

Protective class
I
Overvoltage category / degree of contamination
III / 2
Protective separation
Electrical isolation of the 115 / 230 V AC power supply from
the other current circuits by means of reinforced or double
insulation. Functional extra-low voltages (PELV) on the low
voltage side

CSA certification
The SCC-K NO2/NO converter is certified for use in general
purpose environment, evidenced by full compliance with
standards CAN / CSA-C22.2 No. 61010-1-04 and UL 61010-1
(2nd Edition).
Certificate No. 2298365

Electromagnetic compatibility
Emitted interference
Testing to EN 61000 Part 6-3 : 2001
Immunity to interference
Testing to EN 61000 Part 6-2 : 2005

Ambient conditions
Ambient temperature
Operation: +10 to +50 °C,
Storage and transport: –25 to +65 °C
Relative humidity
 75 % annual mean, occasional and slight condensation
permitted

20

Ordering information
Catalog No.
Converter SCC-K, cock PVDF, 230 V AC
23093-4-0801974
Converter SCC-K, 2 solen. valves, 230 V AC 23093-4-0801975
Converter SCC-K, heated, 1 s.v., 230 V AC 23093-4-0801976
Converter SCC-K, cock PVDF, 120 V AC
23093-4-0801977
Converter SCC-K, 2 solen. valves, 120 V AC 23093-4-0801978
Converter SCC-K, heated, 1 s.v., 120 V AC 23093-4-0801979
Accessories
Wall mounting bracket
23009-4-0801980
Carbon catalyst set
23009-4-0801981

System components and accessories for sample gas conditioning

10/23-5.20 EN Rev. 13

SCC-K NO2 /NO converter
Dimensional drawing (dimensions in mm)
483

1
2

325

132,5 (3HE)

260

44

3
4
5

360
360

NO
NOx

Power supply input X1
Status signal output X2
(9-pin Sub-D female connector)
Heated sample gas inlet 6 mm
Sample gas outlet G 1/4 female thread
Sample gas inlet G 1/4 female thread

465.6

57

120

435

1

2

3

4

5

Circuit and connection diagram
Bei Steuerung nur intern
Brücke X2/1 nach X2/6
Internal controlled only
link X2/1 to X2/6
Converter

Statussignal NO
Status signal NO

X1 F1/2 3,15A

internal
external

X2/5

X2/9

X2/1

X8/4

X8/6

X8/10

Bypass
X2/2

X2/3

X2/6

X8/9

X8/1

X8/8

X9/3

X9/1

X9/4

Netzfilter
mains suppression filter
X3/3

Netz
Power
supply

X3/2

A1 ASP 204M
K1

PE

K2

K3
S1

X3/1
X4/3

X5/4

X5/3

X5/2

X5/5

X5/6
F4

Y1

Y2

+
-

E1

83
142

111

143

112

E1.1

243

X6/5

X6/2
X6/1

L1

80

242

X6/4
X4/2

N

D1 1N4007
T1

230V

81

115V

X6/6

115V

X6/3

D2 1N4007
F3

Frontplatte
front panel

+
C1

D3

470μF
35V

K1

X9/6

D4

X9/5

K2

X5/1

X7/1
0 sw./bl.

X7/2
+ rot/red

M1

10/23-5.20 EN Rev. 13

System components and accessories for sample gas conditioning

21

SCC-K NO2 /NO converter
Pneumatic layouts
Standard version with 4-way ball valve
1 Power supply
2 Status signal
4 Sample gas outlet
5 Sample gas inlet
E1 Tube furnace
Y1 4-way ball valve

4

360
2
E1
1
Y1

5

Option with heated sample gas inlet
1 Power supply
2 Status signal
3 Heated sample gas inlet
4 Sample gas outlet
E1 Tube furnace
Y1 3/2-way solenoid valve

3

360
2
E1
1

Y1

4

4

360
2

Option with two solenoid valves
1 Power supply
2 Status signal
4 Sample gas outlet
5 Sample gas inlet
E1 Tube furnace
Y1 3/2-way solenoid valve
Y2 3/2-way solenoid valve

E1
Y2
1
Y1

5

22

System components and accessories for sample gas conditioning

10/23-5.20 EN Rev. 13

CGO-9 Tubular furnace
Application and description
The tubular furnace is used to convert a non-measurable gas
component into a measurable compound which can be measured by a gas analyzer. This is for example necessary with
certain carbon, sulfur and halogen (such as chlorine) compounds. The conversion is performed in a reaction tube which
is filled with a catalyst and heated by the tubular furnace.
A further application of the tubular furnace is to remove interfering components without the other gas components being
affected.

Technical data
Sample gas pressure
pabs  120 kPa (1.2 bar)
Operating data
Dead time
90% time
in s at
in s at
30 l/h 60 l/h 30 l/h 60 l/h
Without filling 1.5
0.5
2
0.75
With filling
2
0.7
4.5
2.5

Pressure drop
in hPa/mbar at
30 l/h 60 l/h
1.5
3.6
6
12

Dimensional drawing (dimensions in mm)

Reaction tube
Material quartz, volume 20 ml, filling volume 8 ml
Thermocouple
Chromel-Alumel (corresponds to NiCr-Ni)

Weight
Approx. 1.0 kg

Sample gas connection
Glass tube 5 mm outer diameter for hose with 4 mm inner
diameter
Electrical connections
Terminals via cable gland M16
Ambient temperature
Operation: +5 to +50 °C,
Storage and transport: –25 to +65 °C
Power supply
220 V AC +10 to –12 %, 48 to 62 Hz
Power consumption 150 W

3
181

1 Gas inlet
2 Gas outlet (offset by 60°)
3 Mounting bracket or strap

Mounting
Wall fastening

10/23-5.20 EN Rev. 13

4

Z-8514

Degree of protection of housing
IP50 to EN 60529
A protective cap (accessory) is required if the tubular furnace
is not protected against touching by design measures such
as installation in a cabinet or a case.

213

Case material
Steel sheet

5

25

241

Heat-up time
Approx. 120 min.

6
1;2

ca. 120

Furnace temperature
Max. 800 °C
The intermediate temperatures are set using an external
on/off controller.

4 Terminal box
5 Thermocouple
6 Protective cap

Ordering information
CGO-9 with reaction tube,
with thermocouple
CGO-9 with reaction tube,
with catalyst LF 316, with thermocouple
Accessories
CGO-9 reaction tube, quartz
CGO-9 reaction tube, stainless steel with
catalyst LF 316
CGO-9 stainless steel catalyst LF 316

System components and accessories for sample gas conditioning

Catalog No.
23092-4-0806560
23092-4-0856545

23005-4-0806559
23004-4-0856544
23004-4-0801715

23

CGO-9 Tubular furnace
Examples of applications
Application

Furnace
Temperature Reagent
temperature controller

Sample gas Mounting
throughput

Oxidation of sulfur compounds
Combustion of organic chlorine compounds
Conversion of H2 to HCl using Cl2, where
Cl2 concentration  H2 concentration
(O2 concentration > 4 Vol.-%, otherwise use
reaction element, see page 25)
NO2/NO conversion in the exhaust air of nitric
acid or pickling plants.
Combustion of NH3 in air.
Combustion of organic nitrogen compounds,
such as emissions from animal keeping or
carcass incineration.
Application range: 100–3000 ppm NO2

800 °C
800 °C
800 °C

no
yes
yes

none
none
none

60 l/h
60 l/h
60 l/h

750 °C

yes

Stainless steel catalyst LF 316 60 l/h
Service life: change catalyst
every 6 months as a preventative measure.
The stainless steel catalyst is
not suitable for SO2containing combustion gases.

24

System components and accessories for sample gas conditioning

arbitrary
arbitrary
arbitrary

vertical

10/23-5.20 EN Rev. 13

Reaction element
Application and description
The reaction element with UV lamp is used for gases which
react chemically under the influence of UV light. A typical
application is the conversion of hydrogen (H2) with chlorine (Cl2)
into hydrochloric acid (HCl). For chlorine concentrations of less
than 85 Vol.-%, however, the tubular furnace CGO-9 should be
used in principle.
The sample gas is irradiated in a quartz reaction vessel by a
UV lamp. The lamp is mounted in the centre of the reaction
vessel so that the UV light radiates to all sides. This device is
housed in a plastic case. The formation of ozone outside the
reaction vessel is prevented by purging the case with nitrogen
or instrument air.

Technical data
Sample gas pressure
pabs = 80–150 kPa (0.8–1.5 bar)
Dead volume of the reaction tube
Approx. 250 cm3
Service life of the UV lamp
Approx. 5 months

Dimensional drawing (dimensions in mm)

Case material
Glass-fiber-reinforced polyester
Weight
2 kg

6

Degree of protection of housing
IP54 to EN 60529
1

Materials of gas-conducting parts
FPM, quartz

2

Electrical connection
Terminals via PG 11 cable gland

3

Gas connection
Glass tube with 7 mm outer diameter for tube with 6 mm
inner diameter and 9 mm outer diameter

4
5

Case purging
Approx. 10 l/h nitrogen or instrument air
Power supply
230 V AC, 50 Hz. Power consumption 0.5 A during start-up,
0.35 A in continuous operation
1
2
3

Ordering information
Reaction element with UV lamp

10/23-5.20 EN Rev. 13

Purge gas outlet
Sample gas outlet
Power supply

4
5
6

Sample gas inlet
Purge gas inlet
Cable gland

Catalog No.
23095-5-5854124

System components and accessories for sample gas conditioning

25

Membrane filter
Application

Dimensional drawings (dimensions in mm)

Membrane filters are used in gas conditioning systems of
analyzer equipment for fine filtering of dust particles  1 μm, in
order to protect the analysis instruments from contamination in
the event of disturbances in the sample gas conditioning chain.

Membrane filter, PVC version

Technical data
Flow rate (air)
Max. 300 l/h
Gas pressure pabs
PVC: max. 150 kPa (1.5 bar),
stainless steel: max. 200 kPa (2 bar)
Gas temperature
PVC: max. 55 °C, stainless steel: max. 100 °C
Retention rate
100 % for particles  1 μm
Pressure drop
2.5 to 20 hPa (mbar) at a flow rate of 30–250 l/h (air)
Materials of gas-conducting parts
Case: PVC or stainless steel 1.4571 (SAE 316Ti),
gas connections: PTFE, round cord rings: FPM,
filter membrane: glass fiber
Filter surface
Approx. 24 cm2
Dead volume
Approx. 15 cm3
Weight
PVC: approx. 0.59 kg, stainless steel: approx. 1.4 kg

Membrane filter, stainless steel version

Mounting
Panel and wall mounting (with mounting bracket)
Gas connections
G 1/4 female thread or nozzles for hose internal diameter
4 mm
Scope of delivery
Membrane filter, mounting bracket, 2 nozzles, 2 round cord
rings, 50 filter membranes

Ordering information
Catalog No.
Membrane filter
Membrane filter, PVC
Membrane filter, stainless steel
Condensate monitor (see page 27)
Condensate monitor, PVC
Condensate monitor, stainless steel

26

23145-5-8018438
23147-5-8018439
23158-5-8018440
23157-5-8018039

System components and accessories for sample gas conditioning

10/23-5.20 EN Rev. 13

CGKW Condensate monitor
Application
The condensate monitor CGKW is used to indicate moisture
(condensate) or any other electrically-conducting liquid which
penetrates into the sample gas line to the gas analyzer. At the
same time, the condensate monitor is a membrane filter and it
therefore satisfies the double function of fine filtering the sample gas as well as of signaling moisture. The filter membrane
holds back the penetrated concentrate for a short while. An
independent condensate barrier (see application examples) can
be constructed with the condensate monitor as sensor element
and the switching unit ER-144A/Ex (see page 32).

control circuit of the switching unit ER-144A/Ex, the apparatus is
permitted in hazardous areas of Zones 1 and 2 up to a gas and
ambient temperature of 55 °C without restriction of the explosion
groups and temperature classes.

Application examples
The process gas is present at zero pressure or with
negative pressure
Sample gas feeding by pump. Condensate barrier by switching
off the pump. Condensate monitor and analyzer must be
arranged above the sample gas conditioning chain.
7

Technical data
Flow rate (air)
Max. 300 l/h

6
1

2
Z-20002/4

Gas pressure pabs
PVC: max. 150 kPa (1.5 bar),
stainless steel: max. 200 kPa (2 bar)
Gas temperature
Max. 55 °C

5

Retention rate
100 % for particles  1 m
Pressure drop
2.5 to 20 hPa (mbar) at a flow rate of 30–250 l/h (air)
Materials of gas-conducting parts
Case: PVC or stainless steel 1.4571 (SAE 316Ti), gas connections: PTFE, electrodes: stainless steel 1.4305 (SAE 303),
round cord rings: FPM, filter membrane: glass fiber

4

The process gas is present at zero pressure or with weak
positive pressure
Sample gas feeding by pump. Condensate barrier by switching
off the pump and switching over a 3/2-way solenoid valve.
Condensate monitor and analyzer can be arranged as you
wish.
7

Filter surface
Approx. 24 cm2
6

Dead volume
Approx. 15 cm3

1

3

2

Z-20002/5

Dimensions
As for membrane filter (see page 26)
Weight
PVC: approx. 0.71 kg, stainless steel: approx. 1.5 kg
Mounting
Panel and wall mounting (with mounting bracket)
Gas connections
G 1/4 female thread or nozzles for hose inside diameter
4 mm

5

4

The process gas is present with positive pressure
Sample gas feeding by positive pressure. Condensate barrier
by switching over a 3/2-way solenoid valve. Condensate
monitor and analyzer can be arranged as you wish.
7

Electrical connections
Two-core cable, approx. 1 m, for connecting to the switching
unit ER-144A/Ex (see page 32)
6

Mounting orientation
Safety nose at top, electrode below

2

3

Z-20002/6

Scope of delivery
Condensate monitor, mounting bracket, 2 nozzles, 2 round
cord rings, 25 filter membranes, operating manual
5

Ordering information
See page 26

Explosion protection (only with stainl. steel case)
The condensate monitor is a device without voltage source. If
the condensate monitor is being operated in the intrinsically-safe
10/23-5.20 EN Rev. 13

1
2
3
1)

4

Membrane pump
4 Switching unit
7 Analyzer
Condensate monitor
5 Pilot lamp
6 Flow meter
Solenoid valve 1)
The sample gas path to the analyzer is blocked in the
de-energized condition

System components and accessories for sample gas conditioning

27

Acid filter
Application and description

Dimensional drawing (dimensions in mm)

20

The acid filter is used in the gas conditioning system. Its purpose is to collect the sulfuric acid aerosols entrapped in the
sample gas. The acid filter may always be required where the
SO3 content in the sample gas is greater than 1 mg/m3 or the
SO2 content is greater than 1000 mg/m3 – for example in flue
gas from heavy-oil or hard coal firings.

50

6,5

ca. 25

The filter element is made up of borosilicate fibers. An influencing and “hang-up” of the measuring component in the sample
gas is not detectable with this material.
85

Technical data

140

60

Flow rate (air)
Max. 250 l/h
Gas pressure
pabs = 50–200 kPa (0.5–2 bar)
Sample gas temperature
Max. 150 °C
Sample gas dew point
Max. 70 °C
Retention rate
99.99 % for particles  0.1 μm

78
36

DN4/6

23.5

Pressure drop
10 hPa at 250 l/h, 50 hPa at 250 l/h
Dead time
Approx. 20 s
Materials of gas-conducting parts
Connection cap: PVDF; vessel: glass;
filter element: borosilicate glass micro fiber

65

Dead volume
Approx. 130 ml
Dimensions
See dimensional drawing

Ordering information

Weight
Approx. 0.57 kg

Acid filter

Catalog No.
23045-5-8018419

Mounting
Wall mounting with mounting bracket, mounting orientation
vertical
Gas connections
Male fitting (PVDF) for 4/6/1 mm, drainage connection GL14
with terminal fitting 4/6/1 mm (PVDF)
Ambient temperature
+5 to +50 °C
Service life
Change element if there is contamination and perceptibly
higher pressure drop
Scope of delivery
Filter vessel with filter element, two male fittings, drainage
connection GL14, mounting bracket, mounting and replacement instructions

28

System components and accessories for sample gas conditioning

10/23-5.20 EN Rev. 13

Disposable filter
Application and description

Dimensional drawing (dimensions in mm)

The disposable filter is used in the gas conditioning system.
It is mounted in the sample gas line upstream of the gas
analyzer.
The disposable filter consists of a micro fiber filter pipe which is
welded into a plastic case.

Technical data
Flow rate (air)
Max. 250 l/h
Gas pressure
pabs  200 kPa (2 bar)
Sample gas temperature
+5 to +50 °C
Retention rate
99.99 % for particles  0.1 μm

Ordering information

Pressure drop
Approx. 2 hPa at 60 l/h

Disposable filter

Catalog No.
23044-5-8018418

Dead time
Approx. 3 s at 60 l/h
Materials of gas-conducting parts
Case: PA,
Filter element: borosilicate glass micro fiber
Case volume
11.5 cm3
Dimensions
Height 80 mm, diameter 28 mm
Weight
Approx. 0.1 kg
Mounting
Mounting orientation preferably vertical
Gas connections
Two pipe nipples, outside diameter 6.3 mm
Ambient temperature
+5 to +50 °C
Service life
Change filter if there is condensate accumulation (acid), and
in any case every 6 months
Scope of delivery
Disposable filter, mounting and replacement instructions

10/23-5.20 EN Rev. 13

System components and accessories for sample gas conditioning

29

Flow meters, flow monitors
 Flow measurement using float-type flow meters
 Replaceable measuring tube
 Integral needle valve
 Flow monitor with inductive contact

Dimensional drawing (dimensions in mm)

Measuring range
7–70 l/h nitrogen at 20 °C and 1013 hPa (mbar)
The following flow rates are obtained with other gases under
the same conditions at the 60 l/h mark (approximate values):
Sulfur dioxide
SO2 49 l/h
Carbon dioxide
CO2 55 l/h
Argon
Ar
49 l/h
59 l/h
Air
54 l/h
Oxygen
O2
Carbon monoxide CO
60 l/h
Methane
CH4 86 l/h
159 l/h
Hydrogen
H2
Helium
He
77 l/h

42
36

90

111

2

45

Flow meter 7–70 l/h

Gas inlet conditions
Pressure max. 10 MPa (10 bar), temperature 100 °C
Flow rate adjustment
With integral needle valve

~82

Materials of gas-conducting parts
Measuring tube
Borosilicate glass
Float
Stainless steel 1.4401 (SAE 316)
Float catcher
PTFE
Fitting
Stainless steel 1.4571 (SAE 316Ti)
Needle valve
Stainless steel 1.4571 (SAE 316Ti)
Seals
FPM
Nozzles
PTFE

Z-20249

Measuring deviation
5 % of the full-scale value

1

1 Gas inlet
2 Gas outlet

Weight
Approx. 0.5 kg
Mounting
Panel or wall mounting by means of mounting bracket
Gas connections
1/4 NPT female thread or sleeve for hose 4 mm inner
diameter
Scope of delivery
Flow meter, mounting bracket, 2 nozzles

30

System components and accessories for sample gas conditioning

10/23-5.20 EN Rev. 13

Flow meters, flow monitors
Flow monitor 1.6–16 l/h, 7–70 l/h

Dimensional drawing (dimensions in mm)

Measuring range
1.6–16 l/h or 7–70 l/h nitrogen at 20 °C and 1013 hPa (mbar)

2

Measuring deviation
5 % of the full-scale value

~145

Weight
Approx. 0.8 kg

60

Materials of gas-conducting parts
As with flow meter 7–70 l/h

90

Flow rate adjustment
With integral needle valve

Mounting
As with flow meter 7–70 l/h
Gas connections
As with flow meter 7–70 l/h
Signaling
With inductive contact RC 10-14-NO
Float catcher set to 5 l/h or 20 l/h
Electrical connections
Terminal boxes with EMC filter and cable gland M16
Scope of delivery
Flow monitor, mounting bracket, 2 nozzles, O-ring

Flow monitor for chlorine 7–70 l/h

60

1 Gas inlet
2 Gas outlet

Ordering information

Measuring range
7–70 l/h chlorine at 20 °C and 1013 hPa (mbar)
Measuring deviation
5 % of the full-scale value
Gas inlet conditions
Pressure max. 4 MPa (4 bar), temperature 100 °C

Z-20250

1

Flow meter 7–70 l/h
Flow monitor 1.6–16 l/h
Flow monitor 7–70 l/h
Flow monitor for chlorine 7–70 l/h

Catalog No.
23151-5-8018474
23155-5-8018476
23155-5-8018475
23155-5-8018477

Flow rate adjustment
With integral needle valve PEEK
Materials of gas-conducting parts
Measuring tube
Borosilicate glass
Float
PEEK with lining
Float catcher
PTFE
Case
PVDF
Seals
FFKM
Weight
Approx. 0.5 kg
Mounting
Panel or wall mounting by means of mounting bracket
Gas connections
1/4 NPT female thread
Signaling
With inductive contact RC 10-14-NO
Float catcher set to 20 l/h
Electrical connections
Terminal boxes with EMC filter and cable gland M16
Scope of delivery
Flow monitor, mounting bracket, 2 nozzles, O-ring

10/23-5.20 EN Rev. 13

System components and accessories for sample gas conditioning

31

ER-144A/Ex Switching unit
Application and description

Dimensional drawing (dimensions in mm)

The switching unit ER-144A/Ex acts as switching amplifier for
the condensate monitor CGKW (see page 27). The intrinsicallysafe circuit of the switching unit ER-144A/Ex enables the
condensate monitor also to be installed in hazardous areas.
The ER-144A/Ex switching unit operates in closed-circuit
operation. The alarm positions of the output contacts correspond to those of the device in a non-voltage state or exceeding the adjustable measurement current value (electric conductance) or disconnection of the sensor lead. In the “Ready”
status, the output contacts have changed over to the corresponding counter-position (internal relay picked-up).

144/A

99

A 100-k resistor (1 % / 0.25 W) must be installed at the sensor
in order to monitor the signal line between the sensor and the
electrode relay. Without the resistor, the switching unit will
change to the “Line break” position.

Technical data

14 11 N L1

Indication
One LED each for line breakage and condensate penetration

NO2

12 11 14 22 21 24
COM2

21 22 24 12

NO1

Output
2 potential-free changeover contacts, max. 250 V AC, 5 A

Electrical connections

NC2

“ib”
 13.1 V
 5 mA
 65 mW
 6 μF
1H

NC1

“ia”
 13.1 V
 5 mA
 65 mW
 0.97 μF
 0.9 H

COM1

Input
No-load voltage U0
Short-circuit current I0
Power P0
Permissible external capacitance C0
Permissible external inductance L0

Construction
Surface-mounting case
Degree of protection
Housing IP40, terminals IP20 to EN 60529
Electrical safety
Protection class II,
Over-voltage category III,
Degree of pollution 2

E1 E2 E1 E2

Weight
Approx. 0.2 kg
Mounting
Snap fastening on standard rail 35 mm (EN 50022)
Ambient temperature
–25 to +60 °C

Relays are shown in the unpowered state.

Power supply
230 V AC or 115 V AC 10 %, 48–62 Hz,  1 VA
EC type examination certificate
TÜV 00 ATEX 1604
Designation
II (1) G [EEx ia] IIC

Ordering information
Switching unit ER-144A/Ex, 230 V AC
Switching unit ER-144A/Ex, 115 V AC

Catalog No.
23371-4-0730637
23371-4-0730638

Installation location
Outside the hazardous area

32

System components and accessories for sample gas conditioning

10/23-5.20 EN Rev. 13

KFA Isolated switch amplifier
Application and description
The KFA isolated switch amplifier is used as switch amplifier
for the flow monitor (see page 30).
The control circuit is monitored for line breakage and short
circuit. External faults are indicated by a red flashing LED.

Technical data
Input (intrinsically safe)
U0  10.6 V
I0  19.1 mA
P0  51 mW
Permissible connection values
Explosion group
IIA
External capacitance
72 μF
External inductance
780 mH

IIB
16.2 μF
390 mH

IIC
2.32 μF
97 mH

Output relays
Potential-free changeover contacts, max. 253 V AC, 2 A;
cos  > 0.7; switching frequency < 10 Hz
Construction
Surface-mounting case

Electrical connections

Degree of protection of housing
IP30 to EN 60529
Weight
approx. 150 g

Input I EEx ia IIC

Input II EEx ia IIC

1+

4+

1+

1+

EC type examination certificate
PTB 00 ATEX 2081

Installation location
Outside the hazardous area

Switch S3 in position I

10 kΩ

4+

10 kΩ

2+

610 kΩ

3-

without
LB, SC 1+

Power supply
207–253 V AC or 103.5–126 V AC, 45–65 Hz,  1 VA

Designation
II (1) G D [EEx ia] llC

3-

Switch S3 in position I
without SC

Ambient temperature
–20 to +60 °C

10 kΩ

2+

5+

3-

4+

3-

4+

6-

5+

6-

LB
SC
I

II

S1

Switch S3 in position II

6-

LB
SC
S3

red

I

II

S2

&

S3

red
+ -

&

1

Safe area

Mounting
Panel or wall mounting with screw fastening (EN 50022/3) or
snap fastening on standard rail 35 mm (EN 50022)

6400 Ω ≤ R ≤ 2 kΩ

Hazardous area,
zone 0, zone 1

Switch S3 in position I
3400 Ω ≤ R ≤ 2 kΩ

green
yellow

yellow
+

7 8 9
Output I

10 11 12
Output II

14 15
Power
supply

-

Power LB/SC
Rail collective
error message
only
KFD2-SR2-Ex2.W

Ordering information
KFA6-SR2-Ex1.W,
KFA6-SR2-Ex2.W,
KFA5-SR2-Ex1.W,
KFA5-SR2-Ex2.W,

10/23-5.20 EN Rev. 13

1 channel, 230 V AC
2 channels, 230 V AC
1 channel, 115 V AC
2 channels, 115 V AC

System components and accessories for sample gas conditioning

Catalog No.
23372-5-8328644
23372-5-8328645
23372-5-8328851
23372-5-0730651

33

Stripper for cooling water and waste water
Application and description
The stripper enables traces of organic solvents or pollution in
waste water or cooling water to be monitored continuously.
The stripper is used in combination with a flame-ionization
detector (FID). The device operates on the flow principle. 100–
300 l/h of the water to be analyzed are sent through a glass
body in the center of which is located a glass frit which is
continuously purged with the instrument air. Hydrocarbons
change into the gas phase and are driven out of the medium by
the air bubbles. The air stream is then analyzed with an FID.
The stripper unit is operated on the bypass principle without a
pump. The cross-sections of the entire unit are designed in
such a way that clogging is prevented to a greater or lesser
extent. A threefold safeguard against flooding and water breakthrough to the FID is achieved by an overflow and an
emergency overflow as well as a check valve. The feed can be
shut off by means of a gate valve. To obtain long-term
measurement stability, the strip air is fed via flow regulators.

Ambient temperature
+10 to +50 °C

Technical data
Instrument air
100–700 kPa (1–7 bar), consumption approx. 100 l/h

Power supply
230 V AC, 50/60 Hz or 120 V AC, 60 Hz

Minimum pressure difference
10 kPa (0.1 bar)

Ordering information

Medium temperature
+10 to +50 °C
Flow rate sample water
100–300 l/h
Heat-up time
< 7 s with 3 m feed line between stripper and FID

Stripper IP65, unheated
Stripper IP65, heated, 230 V AC
Stripper IP65, heated, 120 V AC

Catalog No.
23841-5-4560001
23841-5-0768362
23841-5-0768361

Materials of gas-conducting parts
Rigid PVC, glass
Case material
G.R.P.
Degree of protection of housing
IP65 to EN 60529
Dimensions
1000 x 460 x 340 mm (H x W x D)
Weight
24 kg
Connections
Connection, test water:
1 inch thread connection
Connection, sample water:
DN 25 using DIN 4-hole flange with flat seal
Discharge, sample water:
Discharge point DN 50 via DIN 4-hole flange with flat seal
Contacts
Potential-free contact for flow monitoring of the sample
water

34

System components and accessories for sample gas conditioning

10/23-5.20 EN Rev. 13

Multi-way test gas cocks
Multi-way test gas cock, PVDF version
Flow rate
Approx. 1000 l/h at 100 hPa pressure difference and 20 °C
Gas pressure
pabs  50–200 kPa (0.5–2 bar)
Leakage rate
 5 x 10–3 hPa I/s
Materials of gas-conducting parts
PVDF, PTFE, FPM
Weight
Approx. 0.7 kg
Mounting
Wall or panel mounting
Gas connections
G 1/4
Ambient temperature
+5 to +80 °C

Multi-way test gas cock, stainless steel version

Dimensional drawing (dimensions in mm)

Flow rate
Approx. 1000 l/h at 50 hPa pressure difference and 20 °C

Leakage rate
 5 x 10–3 hPa l/s
Materials of gas-conducting parts
PTFE, stainless steel SS 316, suitable for pure oxygen

94

62,5
~110

~124

Gas pressure
pabs  17.6 MPa (176 bar)

75

Weight
Approx. 0.8 kg

82

Mounting
Wall mounting (with accessories) or panel mounting

Nullgas
Zero gas

Gas connections
1/8 NPT and screw fitting for pipes with 6 mm outer diameter
Ambient temperature
+10 to +65 °C

Messgas
Meas.gas

Endgas 1
Span gas

83

Ordering information
Catalog No.
Test gas cock PVDF
23177-4-0839859
Test gas cock stainless steel for wall mount. 23177-5-8308581

10/23-5.20 EN Rev. 13

Montagebohrung
Bore holes

45°

System components and accessories for sample gas conditioning

63

M6

35

Cylinder pressure reducers
Application and description
Cylinder pressure reducers are used to reduce the filling
pressure in pure gas and test gas cylinders to the required
operating pressure. Material compatibility must be observed in
respect of the test gas (see technical data). Two-stage cylinder
pressure reducers are employed where very constant pressure
and flow are required at low operating pressure – for example
where the subsequent gas paths are switched over using
solenoid valves. Pressure reducers with minimum contact at
the high-pressure gauge give an indication when the cylinder
pressure has dropped to such a level that it is necessary to
replace the gas cylinder.

Ordering information
2-stage for nitrogen
2-stage for non-corrosive test gases
2-stage for corrosive test gases
1-stage for nitrogen, brass
1-stage for non-corrosive test gases, brass
2-stage for hydrogen, brass
2-stage for hydrogen, stainless steel

Catalog No.
23422-5-8018376
23422-5-8018377
23422-5-8018378
23422-4-0730654
23422-4-0730655
23422-4-0730652
23422-4-0730653

Technical data
Catalog No.
Gas type
Construction
Inlet pressure
Outlet pressure (adjustable)
Safety valve set to
Relief valve set to
Pressure gauge indication
Inlet pressure
Outlet pressure
Flow rate
Control action
Status signal
Materials of gas-conducting parts
Membranes
Seal
Case
Dimensions (W x H x D)
Weight
Connections
Connection to cylinder
(thread to DIN 477)
Gas outlet
Safety valve
Relief valve
Ambient temperature
Scope of delivery

Dimensional drawing
36

23422-5-8018376
23422-5-8018377
Nitrogen
Non-corr. test gases
2-stage, brass, with minimum contact
max. 21 MPa (210 bar)
10 to 150 kPa (0.1 to 1.5 bar)
approx. 1.6 MPa (16 bar)
approx. 200 kPa ( 2 bar)

23422-5-8018378
Corrosive test gases
2-stage, stainless steel, with min. contact
max. 21 MPa (210 bar)
10 to 150 kPa (0.1 to 1.5 bar)
approx. 1.6 MPa (16 bar)
approx. 200 kPa (2 bar)

0 to 25 MPa (0 to 250 bar)
–100 to +300 kPa (–1 to +3 bar)
25 to 500 l/h
Back pressure increase with falling cylinder pressure down to 3 bar equals zero
Reed contact 100 V, 0.5 A

0 to 25 MPa (0 to 250 bar)
–100 to +300 kPa (–1 to +3 bar)
25 to 500 l/h
Back pressure increase with falling cylinder pressure down to 3 bar equals zero
Reed contact 100 V, 0.5 A

Stainless steel
ETFE, PVDF, EPDM
Brass chromium plated
approx. 160 x 185 x 167 mm
approx. 2.1 kg

Stainless steel
ETFE, PVDF, EPDM
Stainless steel 1.4401 (SAE 316)
approx. 160 x 185 x 167 mm
approx. 2.1 kg

Hexagon

Hexagon

Clamping ring screw fitting Swagelok®
for pipe with 6 mm outside diameter
Clamping ring screw fitting Swagelok®
for pipe with 6 mm outside diameter
1/4 NPT male thread
–5 to +75 °C
Pressure reducer, pressure gauge, safety
valve and clamping ring screw fitting
attached; replacement seals, manual
see above

Clamping ring screw fitting Swagelok®
for pipe with 6 mm outside diameter
Clamping ring screw fitting Swagelok®
for pipe with 6 mm outside diameter
1/4 NPT male thread
–5 to +75 °C
Pressure reducer, pressure gauge, safety
valve and clamping ring screw fitting
attached; replacement seals, manual
see above

System components and accessories for sample gas conditioning

10/23-5.20 EN Rev. 13

Cylinder pressure reducers
Technical data
Catalog No.
Gas type
Construction
Inlet pressure
Control range
Control action

23422-4-0730654
Nitrogen
1-stage, brass
21 MPa (210 bar)
0 to 250 kPa (2.5 bar)


23422-4-0730655
Non-corrosive test gas
1-stage, brass
21 MPa (210 bar)
0 to 250 kPa (2.5 bar)


31.5 MPa (315 bar)


23422-4-0730652
Hydrogen
2-stage, brass
21 MPa (210 bar)
0 to 350 kPa (3.5 bar)
Back pressure increase
with falling cylinder
pressure down to 3 bar
equals zero

25 MPa (250 bar)

23422-4-0730653
Hydrogen
2-stage, stainless steel
21 MPa (210 bar)
0 to 350 kPa (3.5 bar)
Back pressure increase
with falling cylinder
pressure down to 3 bar
equals zero

25 MPa (250 bar)

Pressure gauge
Reed contact
pressure gauge
Contact
Outlet pressure gauge
indication range
Materials of
gas-conducting parts
Body

31.5 MPa (315 bar)


0 to +400 kPa
(0 to +4 bar)


0 to +400 kPa
(0 to +4 bar)

falling opening
–100 to +500 kPa
(–1 to +5 bar)

falling opening
–100 to +500 kPa
(–1 to +5 bar)

Brass

Brass

Brass

NBR
NBR

100 μm

DIN 477, No. 10
1
–5 to +75 °C
2.1 kg

NBR
NBR

100 μm

DIN 477, No. 14
1
–5 to +75 °C
2.1 kg

Stainless steel
NBR
PVDF/PTFE
50 μm
3m
DIN 477, No. 1
2
–5 to +75 °C
2.1 kg
Outlet with throttle
approx. 40 l/h and
6 mm screwed fitting

Stainless steel 1.4404
(SAE 316L)
Stainless steel
EPDM
PVDF/PTFE
50 μm
3m
DIN 477, No. 1
2
–5 to +75 °C
2.1 kg
Outlet with throttle
approx. 40 l/h and
6 mm screwed fitting

Membrane
Case seal
Seat
Filter
Connection cable
Bottle connection
Safety relief valve
Ambient temperature
Weight
Special features

Dimensional drawings

10/23-5.20 EN Rev. 13

System components and accessories for sample gas conditioning

37

ABB Limited
Process Automation
Oldends Lane
GL10 3TA Stonehouse
Gloucestershire, United Kingdom
Phone: +44 1 453 826661
Fax:
+44 1 453 829671

ABB Australia Pty Limited
Process Automation
Bapaume Road
2170 Moorebank
New South Wales, Australia
Phone: +61 2 9821 0968
Fax:
+61 2 9400 7050

ABB Pte. Ltd.
Process Automation
2 Ayer Rajah Crescent
139935 Singapore, Singapore
Phone: +65 6773 5961
Fax:
+65 6778 0222

ABB Ltd.
Process Automation
14 Mathura Road
121003 Faridabad, Haryana, India
Phone: +91 129 2279627
Fax:
+91 129 2279692

ABB Engineering Ltd.
Process Automation
10 Jiuxianqiao Lu
100015 Beijing, China
Phone: +86 10 84566688 Ext. 6217
Fax:
+86 10 84567650

ABB Automation GmbH
Process Automation
Stierstaedter Strasse 5
60488 Frankfurt am Main, Germany
Fax:
+49 69 7930-4566
E-mail: [email protected]

ABB Inc.
Process Automation
3700 W Sam Houston Parkway South,
Suite 600, Houston, TX 77042, USA
Phone: +1 713 587 8000
www.abb.com/analytical

Note
We reserve the right to make technical changes or
modify the contents of this document without prior
notice. With regard to purchase orders, the agreed
particulars shall prevail. ABB does not accept any
responsibility whatsoever for potential errors or
possible lack of information in this document.
We reserve all rights in this document and in the
subject matter and illustrations contained therein.
Any reproduction, disclosure to third parties or
utilization of its contents – in whole or in parts – is
forbidden without prior written consent of ABB.
Copyright© 2013 ABB
All rights reserved

Sales

Service

10/23-5.20 EN Rev. 13 02.2012