The Model 113 is a Gas Analyzer that can be built with a variety of sensorsThermal Conductivity- binary gases: H2., He, Ar,in air Fumigants in air (ppm to 100%)Combustible Gas sensor- combustibles (0-100%Electrochemical Sensors- 1-3 sensors: H2S (ppm), O2 (%), SO2(ppm), many others...

Comments

Content

Model 113 EC+ or CG
or TCD Gas Analyzer
Introduction

The Model 113 can have electrochemical sensors such as O2, CO2,
NH3, HCl, H2S & ETO. Up to 2 additional electrochemical sensors for ppm
or % range can be added.This Model 113 can be used for safety monitoring around laboratories, manufacturing operations,chemical or petrochemical plants. This handheld analyzer is robust, weighs only 1.5 pounds,
and is extremely easy to use or calibrate. It is an inexpensive but flexible
safety analyzer. There is storage for 7000 data points. Software for downloading to a PC is included.The Analyzer will operate for > 10 hours on a
rechargeable Ni metal hydride battery.. It can also have a TCD or CG.
Model 113 EC+
Theory of Operation
Electrochemical
Two and three electrode systems are used for
the electrochemical sensors. Sensor technology includes potentiometric (CO2, NH3), fuel
cell (O 2) and amperometric (SO 2, H 2S, ...)
where the sample is oxidized or reduced. Sensors are chosen to maximize prformance and
lifetime. The sensors can be easily interchanged via the sensor board and embedded software.
Controls for the Model 113
Ent-enter and change sensor
On/Off- Battery power
Incr-Function ON, scrolling menu up,
increase number
Decr-Function OFF, scrolling menu
down, decrease number
Bkl-Turns backlight on/off
Menus

Site # 1-7,000
Setup- Setup Auto; Ave. time sec., samples/
hour
Exit- Return to Run
Cal
Cal- Performs zero, set cal value, calibration,
sensor for each channel
Cal Gas- Select name of cal gas
Resp as- Once the 113 is calibrated-change
to direct reading on any of > 15 EC sensors
Alarm- Set Alarm value for audible alarm
Exit- Return to Run mode
Datalogging
The 113 has manual or automatic datalogging
capability for >7,000 points. The software for
data logging is included with the Model 13. It
uses Windows Hyperterminal. The data is
shown as follows:
Data From Hyperterminal
Site Date Time
ppm
495 5/12/2004 15:02:27 171.7
496 5/12/2004 15:03:27 171.6
497 5/12/2004 15:04:27 171.6
End Of Log Data

Log
Manual-Set site #, and manually log each pt. This data can be imported directly into MS
Auto- Set, ave. time (sec) and samples/hr. to
ExcelR
Auto log
PID Analyzers, LLC

OPTIONS
SPECIFICATIONS
· Shoulder carry bag
· Single unit construction
· Hard carrying case
· 8.0" L x 3" W x 2.25"D
· Calibration kit
· Weight 1.5 pounds
· 12 VDC charger with cigarette lighter adapter
· Simple 5 button operation
· Sensors choose from >17 electrochemical
· No keyboard-just keypad
· Combustible gas or TCD are optional
· “Resp as” to setup for direct reading for
changing sensors (gases)
· Alphanumeric display for compound,
Applications:
detector, alarm,range, & logging
Safety, Industrial Hygiene
· LCD digital display that is backlight
Leak detection (H2, He, det. limit 10 ppm)
selectable
· Fast response
Features
· Datalogging for> 7,000 points
· 4 Sensor inputs (1-3 EC sensors plus CG) • Easy to use
· Low battery indicator & automatic shutdown • Easy to calibrate
· RS232 output
• Up to 3 EC sensors
· Analog output for recorder
• Fast response- Instant on electrochemical sen
· Simple pushbutton sensitivity control
sors
Alphanumeric scrolling
Upper numeric display is the concentration: • Datalogging for 7,000 points
the lower scrolling display provides
information on the type of sensor, and the
status of logging and alarms.

Sensor

Specifications of Gas Sensors for the Model 1113
Range ppm
Det. Limit
Response Time
Interferences

Electrochemical
Ammonia
Carbon Dioxide
Carbon Monoxide
Chlorine
Ethylene
Ethylene Oxide
Formaldehyde
Hydrogen
Hydrogen Cyanide
Hydrogen Chloride
Hydrogen Sulfide
Nitric oxide
Nitrogen dioxide
Oxygen
Ozone
Phosphine
Sulfur dioxide
Silane

0-50
0-1,000/0-10%
0-500/1000
0-10, 0-100
0-20
0-20
0-30
0-1000
0-50
0-100
0-100
0-50
0-10
0-30%
0-10 ppm
0-5
0-20
0-5

0.1
ppm
0.5
0.1
0,1
0.1
0.1
2
0.1
0.1
0.01
0.1
0- 0.1
0.1%
0.2
0.05
0.1
0.05

30 sec.
90 sec.
10 sec.
30 sec.
45 sec
1 min.
50 sec.
45 sec.
50 sec.
1.5 min.
20 sec.
10 sec.
15 sec.
15 sec.
75 sec
40 sec.
20 sec.
30 sec.

acid gases
other acid gases
C2H4
——
alcohols, other org.
alcohols, org., CO
alcohols, CO
CO, C2H4
C2H4, H2S, SO2
SO2 ,H2S---——
NO2
Cl2, H2S
——
____
SiH4, GeH4, B2H6
NO2
AsH3, GeH4, PH3

There are more than20 sensors available for the Model 1 3. One EC sensoror any three additional EC
sensors or CG or TCD can be incorporated into the head.

PID Analyzers, LLC
2 Washington Circle
Sandwich, MA 02563 USA
T 1 774 413 5281; F 1 774 413 5298

Representative

PID Analyzers
Model 113 Gas Analyzer/ http://www.hnu.com

Sponsor Documents

Or use your account on DocShare.tips

Hide

Forgot your password?

Or register your new account on DocShare.tips

Hide

Lost your password? Please enter your email address. You will receive a link to create a new password.

Back to log-in

Close